Exsenco U.S Contact For Santest

 

GYDC-05

Photograph 11
The GYDC-05 controller position of the magnet digital output (parallel) of resolution 1 millimicron, is the new model controller which achieves the analog output of position or speed (option) simultaneously. Also SSI output and incremental output are possible with option. Because the DIN rail is utilized in the installation of the product, installation and the removal are easy.Combining, as for the probe which you can use GYcRS, GYMR5, GYFRS and GYGS
It is GYPM, GYHR, GYcRP and GYMR4.
11
11
 
11

Specifications

11
  Standard Option
Digital position output
Phase connector (the Sumitomo 3M corporation make)
10136-3000PE attaching
Parallel
Binary (with user to Gray modification possibility)
Negative logic (at the time of TrON output “1”)
Incremental output addition
SSI output addition
Resolution: 0.01mm/bit
(It combines, probe GYcRS, MR5 and FRS)
0.1mm, 0.05mm and 0.002mm
0.001mm/bit
Resolution: 0.1mm/bit
(It combines, probe GYGS, PM, HR, cRP and MR4)
-
Analog position output
(OUT1)
0 - 10VDC
(Load current Max5mA, load resistance Min2kΩ)
4 - 20mA (load resistance Max500Ω) and the like
Resolution: 16bit (1/65536)
(It combines, probe GYcRS, MR5 and FRS)
Resolution: 0.01%FS
(It combines, probe GYGS, PM, HR, cRP and MR4)
Analog speed output
(OUT2) (Note 1)
- ±10V or 4 - 20mA
Supply power source +24VDC±5% (below 160mA)
Alarm output Open collector 0.1A 30VDC (cable broken wire and magnetic end)
Scanning frequency 1kHz (to stroke 1000mm)
Temperature characteristic ±10ppmFS/℃
Category temperature range 0 - +65℃
Storage temperature range -20 - +75℃
* The above-mentioned precision is applied to the sensor above effective stroke 300mm.
(Note 1) in addition to position output, it is speed output additional option.
±10V: At the time of magnetic stop to 0V and probe point when moving +10V.
4 - 20mA: At the time of magnetic stop regardless of to 4mA and portable direction 20mA.
11

Type Variations

11
 
GYDC-05□-□□-□-□-Z□-□/□-□
1 Digital output
1: Standard (parallel output and negative logic binary)
2: Incremental it comes out and regains one's strength (250kHz)
5A: Asynchronous SSI coming out regaining one's strength
5S: Synchronous SSI output being attached
2 Resolution
D2: 0.1mm
D3: 0.05mm
D4: 0.01mm (standard)
D7: 0.002mm
D8: 0.001mm
3 Output direction
D: Magnet to point when moving increase
R: The magnet to the point when moving decrease (when there is no appointment, @AB 1-D4D [or 1-D2D] it becomes)
4 Probe type
11
Sign Probe type Sign Probe type
RS GYcRS GS GYGS (as for resolution only D2)
R5 GYMR5 PM GYPM (as for resolution only D2)
FS GYFRS HR GYHR (as for resolution only D2)
RP GYcRP (as for resolution only D2)
R4 GYMR4 (as for resolution only D2)
5 Effective stroke (mm)
11 You follow largest effective stroke probe specification.
6 Basis dead zone size (mm)
11 You follow basis dead zone size probe specification.
7 Analog position output (OUT1)
11
Sign Output
AD 0 - 10V (standard)
AR 10 - 0V
BD 4 - 20mA
BR 20 - 4mA
CD or CR□□
Bipolar output
(Example)
CD10
CD05
CR10
□□V - □□V


-10V - +10V
-5V - +5V
+10V - - 10V
VZ/F
(Example)
V1/5
V0.5/9.5
Designated value voltage

1 - 5V
0.5V - 9.5V
IZ/F
(Example)
I5.12/20
I20/5.38
Designated value electric current

5.12 - 20mA
20 - 5.38mA

* Z= zero point (basis) output F= full scale point (point) output
8 Option: Analog output (OUT2)
In case of position and output, it selects from F
Speed in case (Note 1) of output
VA [] ±10V
WB [] 4 - 20mA
   []: Maximum path velocity (1.00 - 999mm/sec) < 3 column significant digits, as for decimal point with R inscription >
N: Option unnecessary
9 Magnetic or float form
11
Sign Magnetic type Sign Float type
M0 No.0 F28S φ28SUS316
M0SM No.0SPM F30S φ30SUS316L
M0LM No.0LPM F40S φ40SUS316(B)
M2P No.2P F42S φ42.5 sphere SUS316
M2PN No.2PN F50S φ50SUS316
M3 No.3 F54S φ54SUS304
M11 No.11 F25N RF-A10 plastic float
M11N No.11N F28N RF-A6 plastic float
T142 No.T14-M2 FL□ Float other than describing
T144 No.T14-M4
T162 No.T16-M2
T163 No.T16-M3
MG□ Magnet other than describing
(Note 1)
11 VA at the time of magnetic stop when moving +10V is to 0V and the probe point.
WB at the time of magnetic stop is 20mA regardless of to 4mA and portable direction.
11

External Dimensions Drawing

11
11 ●GYDC-05 controller 11

10